Introduce

    SLP-1000 is able to measure stress distribution of chemically tempered glass which is strengthened by ion-exchange from Li+ to Na+ using scattered light photoelectricity. In case of that the glass is strengthened by mixed liquid KNO3 and NaNO3, K+ layer should be measured by FSM6000 and Na+ layer should be measured by SLP-1000. These data can be combined by specialized software.

*The combination requires optional FsmV dongle

Feature

Model Wavelength CT-CV DOL-Zero
SLP-1000 640nm 5.65Mpa 2.16㎛
SLP-2000 520nm 1.51Mpa 1.42㎛
SLP-2000 405nm 1.00Mpa 1.27㎛
  • Actual data measuring the standard glass 20 times
  • Required Refractive index and photoblastic constant at the wavelength for measurement

Specification

Meas Range CS 0-2000Mpa, DOL 10-400㎛
Meas Resolution Stress 5Mpa    Depth 5㎛
Meas Precision  50㎛ or deeper from surface   Stress± 10Mpa   Depth±10㎛
(For standard glass)
Light source LD Wavelength 640±10nm  30mw   Class 3B
Application Chemically tempered glass, DIOX glass
  Thermally tempered glass
Sample shape Flat-1000R   10x10mm or more
Prism nD=1.516
PC Pre-installed OS, specialized software
OS Windows 10 professional edition 
Size(main body) 320x280x220mm
Weight( main body) 10kg