Introduce
- Recently, luminance and chromaticity management of FDD, Automotive instruments, Fluorescent substance and optical device has been increased. BM-5AC is a high sensitivity luminance colorimeter combines high reliability, operability and high speed.
Feature
- 5 steps selectable measurement angle.
- Built-in analog output, it can be data collect a data with a oscilloscope.
- High speed response speed 0.05ms (Max).
It can measure the object from particle area to wide area without attachment lens.
Building up and falling down time of the flicker light can be measured using Oscilloscope.
Application
- Measurement luminance and chromaticity.
– TV, PC, Tablet, Smart phone display. – Ultra micro spot for automobile interior panel. – Fluorescent material, phosphorescence material, special light source.
Specification
Photo detector | Photomultiplier tube |
---|---|
Optical system | Objective lens : f=80mm F2.5, Eyepiece lens : View field 5° |
Spectral sensitivity | Similar to CIE1931 color matching function |
Measurement angle | 3°/2°/1°/0.2°/0.1° *selectable |
Measurement distance | 350mm – ∞ (Distance from metallic tip of objective lens) |
Measurement range | Auto/Manual 5-range |
Function(*1) | x, y, L (x, y: chromatic coordinate, L: luminance) and ±⊿ u’, v’, L (u’, v’: chromatic coordinate, L: luminance) and ±⊿ X, Y, Z (X, Y, Z: Tristimulus value) and ±⊿ Tc, duv, L (Tc: Correlated color temperature and Deviation) and±⊿ CIE1976 L*a*b*, ⊿Eab* and±⊿, CIE1976 L*u*v*, ⊿Euv*and±⊿ |
Measurable range(*2) | 0.00005 – 1,200,000 cd/m² |
Accuracy in luminance(*3) | ±4% or less |
Accuracy in chromaticity | (1) dx,dy : ±0.005 or less (*3) (2) dx,dy : ±0.008 or less (*4) |
Repeatability in luminance(*5) | 2.0% : 0.005 – 0.025cd/m² 0.8% : 0.025 cd/m² or more |
Repeatability in chromaticity(*6) | 0.003 |
Measurement time(*7) | Approx. 2sec |
Response speed | NORMAL MODE (All Range) : Approx. 30ms FAST MODE : Range-1: Approx. 5ms Range-2: Approx. 0.5ms Range-3: Approx. 0.05ms Range-4: Approx. 0.5ms Range-5: Approx. 0.05ms |
Analog output | DC 0 – 4V (X2・Y・Z) |
Display | Dot-matrix 20 characters x 4rows with backlight |
Interface | RS-232C, USB |
Power supply | Dedicated AC adapter AC100V-240V, 50/60Hz |
Power consumption | Approx. 20W |
Use conditions | Temperature : 0-40℃, Humidity : 85% RH or less (No condensation) |
Outer dimensions (W×D×H) | 154×355×212mm |
Weight | 3.6kg |
(*1)±⊿ : difference of standard value. (*2)For standard illuminance A. (*3)For standard illuminance A, Auto Rage. (*4)For a combination of the standard source A and the next colored glass, when COLOR ADJUSTMENT is applied. O-55,Y-48,A-73B,IRA-05,T-44,R-61,B-46,V-44,G-54) (*5)For standard illuminance A, measuring angle 3°, 2σ, Single mode, Auto Range. (*6)For standard illuminance A, 2σ, Single mode, Auto Range. (*7)Single measurement mode.
Measurement area (Diameter φ)
Measurement angle |
Measurement distance [mm] (*) | ||||
---|---|---|---|---|---|
350 | 500 | 1000 | 5000 | 10000 | |
3° | 15.0 | 23.1 | 49.2 | 255 | 510 |
2° | 10.0 | 15.4 | 32.8 | 169 | 341 |
1° | 5.0 | 7.7 | 16.4 | 85 | 170 |
0.2° | 1.0 | 1.5 | 3.3 | 17 | 34 |
0.1° | 0.5 | 0.8 | 1.6 | 8 | 17 |
(*) The measurement distance is the distance from the tip of the metal fixture on the instrument of the objective lens.
Display Luminance Range(cd/ m²)
Measurement angle |
Range-1 | Range-2 | Range-3 | Range-4 | Range-5 |
---|---|---|---|---|---|
3° | 0.00005-0.15 | 0.0005-1.5 | 0.005-15 | 0.05-150 | 0.5-1,500 |
2° | 0.0001-0.3 | 0.001-3 | 0.01-30 | 0.1-300 | 1-3,000 |
1° | 0.0004-1.2 | 0.004-12 | 0.04-120 | 0.4-1,200 | 4-12,000 |
0.2° | 0.01-30 | 0.1-300 | 0.4-1,200 | 10-30,000 | 100-300,000 |
0.1° | 0.04-120 | 0.4-1,200 | 4-120,00 | 100-300,000 | 400-1,200,000 |
Accessories
■ Attachment Lens : AL-6 ■ Attachment Lens : AL-11 ■ Attachment Lens : AL-12 ■ Reference White Board : WS-3 ■ CCD Adapter : IA-2 ■ Fiber probe : FP-3P ■ Tripod : Tripod5N ■ Tripod : Tripod-SR ■ Fine Control Table : S-4 ■ Mesh Filter : MF-10/MF-100