Introduce
- The CMS-35SPC is designed for measuring the reflection and transmittance, equipped with an integrating sphere for the source of diffuse illumination. The double beam design provides precise and stable measurements. The specimen table can be placed vertically or horizontally, allowing fast and easy placement of samples and making much more efficient use of your operator’s time. The CMS-35SPC can measure the samples with remarkably low spectral reflectance or transmittance, making it ideal for the color management of deep colors.
Feature
- CMS-35SPC adopts a large integrating sphere with a diameter of 200 mm and a full-fledged double-beam colorimetry method.
- Spectral reflectance, spectral transmittance, various color values (X, Y, Z, L*, a*, b*, ⊿a*, ⊿b*, ⊿L*, ⊿Eab*, YI, etc.) can be output as measurement parameters.
- The specimen table can be placed vertically or horizontally
- When placed vertically, measurement can be performed simply by placing a sample on it.
- Reflection measurement of liquid is possible when placed horizontally
- Measurement area can be switched (during reflection measurement)
- Can measure the transmissible color
Measurement of reflection specimen by main body vertical placement |
The transmissible specimen set on the holder |
The transmissible specimen set into the specimen chamber |
Application
- Color management of color materials, including dye, pigment and printing ink, others
- Spectral color measurement used in the calculation of the CCM recipe in color mixing industry, including dyeing, painting, and coloring resin
- Investigating color difference in painting, dyeing and coloring resin
- Achievement of color correction information of the batch color comparing to the standard color guide
- Investigation on the reasons that result in metamerism
- Color management of optical films (YI value, etc.)
- Spectral transmittance measurement of plastics and glasses
- Liquid color and spectral transmittance measurement
- Measurement of transmission color and spectral transmittance of other transmissive objects
Specification
Model | CMS-35SPC | ||||||
Design | Double beam | ||||||
Illuminating and Viewing Geometry | For Reflection White light diffuse illumination by φ150mm integrating sphere / Eight-degree geometry, Specular component included (SCI) measurement mode and Specular component included (SCI) mode are selectable. For Transmittance White light diffuse illumination by φ150mm integrating sphere / Normal geometry |
||||||
Measured Area | For Reflection 7.5x15mm(L)/4x8mm(S) Selectable For Transmittance Diameter 10mm |
||||||
Light source | Halogen Lamp 12V/50W | ||||||
Dispersing Element | Concave Diffraction Grating | ||||||
Spectral Range | 390nm-730nm | ||||||
Wavelength Interval | 10nm | ||||||
Detector | Silicon Photodiode Array | ||||||
Power Requirements | AC 100V, 50/60Hz, 3A | ||||||
Dimensions | 270Wx500Dx300Hmm | ||||||
Weight | 20kg |