Introduce

    SR-5AS is a new generation spectroradiometer that can measure from ultra-low luminance to ultra-high luminance with high sensitivity and high speed.

Feature

  • High dynamic range measuring(0.0001 – 500,000,000cd/m²)
  • Reduction of measurement speed. It can measure more speedy using High speed mode.
  • Significant improvement in usability by adopting a color touch panel.
  • Pursuit of the best function and compatibility for measurement system installation.
  • High accuracy measurement of flashing light (Synchronous measurement function / Integral time delay function).
  • Standard accessory software can control SR-5AS and can process measured data with simple operation.

Application

  • Luminance, Chromaticity, Color rendering index and Color temperature for LED Back light, LED lighting.
  • Luminance and Chromaticity of µLED/Mini LED、OLED、LCD, Tablet, Smart phone display.
  • Luminance and Chromaticity of head light and interior panel of automobile.
  • Used as optical characteristic evaluation sensor for system equipment.
  • Production line(high precision gamma measurement with spectral measuring),R&D(various evaluation such as IVL-measuring).
  • Used as optical characteristic evaluation sensor for system equipment.).

Specification

Photo detector  Electronically cooled linear CCD
Dispersion element Diffraction grating
Optical system Object lens : f= 82mm F2.5, Eyepiece lens : 5° view field
Measuring angle 2°/1°/0.2°/0.1°  Selectable (Electric motor moved)
Measuring distance 250mm – ∞ (Distance from metallic tip of objective lens)
Spectral band width 5nm (Half bandwidth)
Spectral accuracy ±0.3nm *Hg emission lamp
Wavelength range 380 – 780nm
Wavelength resolution 1nm
Measurement mode Auto/Manual(Integral time/Frequency)
External vertical sync signal input
Measuring object Spectral radiance : W・sr-1・m-2・nm-1
Measuring function Radiance Le : W・sr-1・m-2, Luminance Lv : cd・m-2,
CIE1931 chromaticity coordinates xy,
CIE1976 chromaticity coordinates u’ v’, Tristimulus value XYZ,
Correlated color temperature and Deviation,
Dominant wavelength : nm , Peak wavelength : nm
CIE standard observer 2°/ 10°
Measuring luminance range(*1)(*2) 2°:0.0001 – 1,500,000 cd/m2

1°:0.0005 – 4,500,000 cd/m2

0.2°:0.0125 – 100,000,000 cd/m2

0.1°: 0.05 – 500,000,000 cd/m2

Accuracy : luminance(*1) ±2% or less
Accuracy : chromaticity:(x,y) (*1) x : ±0.0015、y : ±0.001
(2° : 0.0005 cd/m2 – 、1° : 0.0015 cd/m2 – 、0.2° : 0.0375 cd/m2 – 、0.1° : 0.15 cdm2 – )
Repeat accuracy : luminance(*3) 1.5% : 0.0005 – 0.005 cd/m2
0.4% : 0.005 – 0.1 cd/m2
0.3% : 0.1 cd/m2  –
Repeat accuracy : chromaticity(*4) 0.005 : 0.0005 – 0.005 cd/m2
0.0015 : 0.005 – 0.1 cd/m2
0.0005 : 0.1 cd/m2  –
Polarization characteristics Luminance 1% or less
Spectral radiance 2% or less (400 – 780nm)
Interface RS-232C、 USB 3.0
Power supply Dedicated AC adapter AC100V-240V, 50/60Hz, DC12V
Power Consumption Approx. 30W
Operating conditions Temperature: 5 – 30℃, Humidity: 80%RH or less
(No condensation)
External dimensions (W×D×H) 422×130×244mm
Weight Approx. : 5.5kg

*1 For standard illuminant A.

*2 Measurable range in Normal and High speed mode.

*3 2σ from 10 times continuous measurement at measuring angle 2° in normal speed mode.

*4 (Max – Min) value from 10 times continuous measurement at measuring angle 2°in normal mode.

Measurement diameter: mmφ

Measuring angle Measurement distance [mm]*
250 350 500 800 1000 2000
6.5 10.0 15.1 25.4 32.2 66.4
3.25 4.99 7.55 12.7 16.1 33.2
0.2° 0.65 1.00 1.51 2.54 3.22 6.64
0.1° 0.33 0.50 0.76 1.27 1.61 3.32

* The measurement distance is the distance from the tip of the metal fixture on the instrument of the objective lens.

Accessories

    ■ Attachment Lens : AL-6 ■ Attachment Lens : AL-11 ■ Attachment Lens : AL-12 ■ Reference White Board : WS-3 ■ CCD Adapter : IA-1A ■ Fiber probe : FP-3P ■ Tripod 5N : Tripod5N ■ Fine Control Table : S-4 ■ Neutral Density Filter : ND-10/100 ■ Illuminance Adapter : ZV-30