SPIRAL VISCOMETER(PCU-285)
International standard for solder paste viscosity measurement
Feature
- Measure TI (Thixotropic Index) & R (Viscosity Non-Recovery Rate) according to JIS Z3284.
- Newly-designed Compact Thermostat realizes better workability.
- Touch Panel Display improves usability and visibility.
- Automatic measurement and save data.
- Optional Printer is also available.
Specification
Model | PCU-285 | ||||
Viscosity Sensor | MALCOM Spairal-pump type | ||||
Viscosity Range | 5Pa・s ~ 800Pa・s | ||||
Speed Range (N) | 1~50rpm | ||||
Shear Rate (D) | (0.6 × N)s-1 | ||||
Measurement Accuracy | ±5% indicated value | ||||
Speed Accuracy | ±2% | ||||
Repeatability | ±0.5% | ||||
Temp. Measurement | 0~50℃ | ||||
±0.5℃ | |||||
Temp. Control Range | Room temperature +/-5C (setting range15~35C) with built-in heater unit | ||||
Display | Touch Panel Display 7Inch | ||||
Measure Mode | PC Connect、PCU-200-Mode、Program Measurement-Mode | ||||
Build-in Printer(option) | Temp, Visc, S.Rate, RPM, Date, Time | ||||
(It must be ordered with PCU-285) | |||||
Automatic Measurement | JIS Standard method、Saved Inspection Program(MAX 5Memories) | ||||
Online measurement ※ | |||||
Interface | USB-Type-A×2、USB-Type-B×1, Ethernet×1 | ||||
Power Supply | AC100~240V 50/60Hz | ||||
Power consumption | Max130w | ||||
Usage environment | Temperature: 10 ℃ to 40℃ Relative humidity: 10 to 90% (no condensation) | ||||
Outer Dimension | 275W x 440H x 215D (mm) | ||||
Weight | Approx.13kg |